庫號:M313718
Tester/*
Helicobacter Pylori (Hp) Tester/*
[Product name]
Helicobacter Pylori (Hp) Tester
[Type specification]
HDW-20-1
Pylori (Hp) Tester
[Product structure and components]
The product consists of detector (ionization type), mechanical device, electronic circuit, and gas collection card.
[Applicable scope of product]
It is applied with Carbon [14C] - urea breath test kit in the clinical diagnosis of Helicobacter Pylori (Hp) infection.
[Safe type]
1. It falls into the category of germ analysis system under clinic examination analyzer (6840) when classified as per medical equipment management. The management category is class II.
2. The electric shock absorber prevention class of the apparatus is Class I, II (over-voltage class). The pollution class of the equipment is class II.
[Operating condition]
1. Ambient temperature: 5℃-30℃;
2. Relative humidity: ≤75%;
3. Atmosphere pressure: 75-106KPa;
4. Power Voltage: AC 220V ± 10% (47-51) Hz;
5. Preheat time: ≥10min.
[Storage condition]
1. Scope of Ambient temperature: -10℃-60℃;
2. Relative humidity: ≤75%;
3. Stored in well-ventilated room with no corrosion gas.
[Performance]
1. The efficiency of detection on 14C unquenched standard source is ≥15%;
2. The background count of 14C is ≤40CPM;
3. The tolerance of relative variation of 14C detection efficiency is ≤30% after the operation of the apparatus for successive 48 hours.
[Usage]
1. Start the equipment
2. The Patient blows air
3. Conduct a measurement
The default measurement time of the apparatus is 250s, put the sample to be measured into the measurement room, and then press the RUN key to start a measurement.
4. Make a measurement report
The result is displayed on the apparatus and printed upon the completion of the measurement.
[Cautions, warnings and suggestion]
1. Press the Reset key to restore an initial state to the apparatus in case of malfunction (the malfunction indicator flashes).
2. Do not touch the gas collection card during measurement. No broken gas collection card is allowed to be used.
3. 1 on the power switch indicates power on and 0 indicates power off.
4. No part of the apparatus should be removed by the user.
5. Please keep the operational environment where the apparatus is used clean and conduct regular cleaning of the measurement room.
6. Never expose the apparatus to strong light.
7. It is recommended to conduct an alignment of the apparatus at the hospital at least once a month. For details, see the method mentioned below:
8. Contact the vendor in time in case of any unrecoverable fault.
[FAQ and maintenance]
1. The apparatus cannot perform a measurement
2. The count is beyond normal scope
3. The printer does not work or prints random codes
[Warranty and after-sale service]
1. Warranty card (attached separay)
Please do confirm and fill relevant information in on the warranty card with your local dealer, such as the date on which you purchased this equipment, read the warrant card carefully and keep it well.
2. Make assignment for repair
Make sure the equipment is powered off in case you cannot shoot the trouble, and contact your dealer or vendor.
3. Apparatus in warranty period
Warranty period: within one year from the date on which the user commissions and accepts the apparatus. Repair is conducted for apparatus in warranty period in accordance with relevant rules of the company.
4. Apparatus outside warranty period
For maintenance of apparatus beyond warranty period, it should be conducted in accordance with relevant rules of the company.
5. After-sale service company: midwest group