· Measures Conductivity, Resistivity, % Concentration, TDS, ppm
· 96 X 96 mm Panel Mount Enclosure Takes Up Less Panel Space
· Temperature Slope Adjustment Provides Greater Accuracy in Chemical
Concentration Control
· “Live Check” Continuously Monitors
Sensor Performance
· Up To Four Alarm Contacts Including
Temperature
· Dedicated Fault Alarm
· NAMUR Fault Current Alarm
· PID Controller
· 2nd Current Output for Temperature
Applications
· Ultrapure Water
· Reverse Osmosis
· Demineralizers
· Boiler Water and Cooling Water
· Metal Cleaning and Plating
Description
The Model 223 Conductivity Transmitter with the appropriate sensor measures conductivity
up to 2000 mS/cm, resistivity in megohms and concentration in percent, ppm, mg/l or TDS.
The basic transmitter is available with either a 115 VAC or 24 Volt AC/DC power supply,
linear isolated current output and one alarm. Features include the addition of a second output
for temperature and up to four alarms which may be configured for on-off control functions,
alarm limits, timer for cleaning or PID control functions. The PS version adds the diagnostic
function Live Check as well as a non-linear programmable current output. Live Check signals
an alarm when the sensor signal does not change over a defined period.
Base Transmitter (PR) Plus Package (PS)
Measurement, display and calibration of the process Adaptive Calibration for Inductive Conductivity Cells
variable (PV) and temperature.
Isolated current output (Code 0) Polarization Detection
Temperature output (Code 1)
One alarm (Code 05) Live Check sensor signal monitoring and alarm.
· On – Off
Dedicated fault alarm (Standard) Programmable non-linear current output.
NAMUR fault current (Standard)
· 2.4 or 22.0 mA
Code 10 Additional Features
Two Alarm Contacts for PV and Temperature USP 24 Alarm for Pharmaceutical Grade Water
· On – Off (WFI Systems)
· Timer for cleaning
· P(ID) Controller Automatic Cleaning initiated by an internal alarm or limit
violation
Code 16 Additional Features
Four Alarm Contacts for PV and Temperature Cleaning initiated externally or automatically by an internal
· On – Off alarm or limit violation
· Timer for cleaning
· P(ID) Controller
Specifications
Conductivity Measurement
Measuring Range Conductivity; 0 to 2000 mS/cm
Resistivity; 0 to 20 Megohms cm
Concentration; 0 to 9999 (%, ppm, mg/l, TDS)
Resolution 0.5% of Measured Value
Repeatability 0.2% of Measured Value
Deviation, Output / Input 0.75% of Current Output Range
Compatible Cell Constants 0.0025 to 99.99 cm-1
Measuring Frequency Conductivity; 299.75 to 1077.6 Hz
Resistivity; 32.5 to 425 Hz
Maximum Cable Length Sensor to Transmitter Conductivity; 100 meters (328 ft.)
Resistivity; 20 meters (65 ft.)
Inductive Conductivity Measurement
Measuring Range 0 to 2000 mS/cm (uncompensated)
Cell Constant 0.0025 to 99.99 cm-1
Measuring Frequency 2 kHz
Maximum Cable Length Sensor to Transmitter 55 meters (180 ft.)
Temperature Compensation
Temperature Sensor Pt 100 or Pt1000
Reference Temperature 25oC (77oF)
Measuring Range -35 to 250oC (-31 to 482oF)
Resolution 0.1oC
Repeatability 1.0% of Measuring Range
Compensation Curves Linear, NaCl, Ultrapure Water, Programmable
Digital Inputs 1 and 2
Voltage 10 to 50 V
Power Consumption Maximum 10 mA
Current Output
Range Selectable 0 to 20 mA or 4 to 20 mA
NAMUR Fault Current 2.4 or 22.0 mA
Maximum Load 500 Ohms
Output Scale Expansion 10% to 99%
Output Isolation Max. 350Vrms / 500Vdc
Over Voltage Protection EN 61000-4-5:1995
Auxiliary Voltage Output
Output Voltage 15 V
Output Current Max. 10mA